| Model Number |
9003 B |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the device did not work during the procedure.The device was replaced and there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the physician palpated the valve prior to implant and decided that it was not working properly.The cause of the issue was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.The valve met the requirements for reflux, pressure-flow, pre-implantation, and leak testing.Therefore the conditions of this complaint could not be duplicated by laboratory personnel.Valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|