MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS CAPSUREFIX NOVUS; ELECTRODE, PACEMAKER, PERMANENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS CAPSUREFIX NOVUS; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 407652
Medical Device Problem Codes Over-Sensing (1438); Under-Sensing (1661)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited oversensing and undersensing during episodes of ventricular tachycardia (vt) and non-sustained ventricular tachycardia which occurred during mitral valve repair surgery.All lead measurements were within normal limits at a subsequent interrogation.The lead remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
MDR Report Key7503469
Report Number3008973940-2018-01121
Device Sequence Number13494655
Product Code DTB
UDI-Device Identifier00763000014643
UDI-Public00763000014643
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/02/2019
Device Model Number407652
Device Catalogue Number407652
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2018
Initial Report FDA Received Date05/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
-
-