| Model Number |
LF1520 |
| Medical Device Problem Code |
Energy Output To Patient Tissue Incorrect (1209)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during colectomy end to end anastomosis, the device worked fine in the beginning but 4 hours later, tissue was not sealed.An end tone was heard, there was no bleeding.The device was disconnected from the generator and connected again, but the issue was not solved.New ligasure was opened and it worked fine.The removed handpiece was checked, and the handle, trigger, worked fine and no abnormality was observed.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in various positions to detect any problematic activation issues.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.The instruction for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|