| Catalog Number |
109.140 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
09/09/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.After the evaluation was noticed that the item received is different from the item reported on guarantee form by the dentist.Therefore, the lot number was not considered.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.Moreover, it was seen that the dentist has used sequence of drills different than the ones recommended for the implant installation.
|
| |
|
Event or Problem Description
|
|
(b)(4) - the dentist reported that after 1 year and 3 months dental implant was installed in intraoral region 25#, its non-osseointegration was observed.In addition, 30 ncm of primary stability was achieved and the patient presented bone type iii.
|
| |
|