• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV PHACO FRAGMENTATION NEEDLES, 23G / 0.6 MM, INCL. NEEDLE WRENCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV PHACO FRAGMENTATION NEEDLES, 23G / 0.6 MM, INCL. NEEDLE WRENCH Back to Search Results
Catalog Number 3005.F106
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 07/14/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research (b)(4)).Details of this activity were discussed with (b)(4) office of compliance (b)(4) during a tele-conference on 05/31/2017.Electron microscopic investigation was done by a external company to determine the root cause.The investigation revealed no results that lead to indication how the issue could occur.Since it was reported small metal pieces coming from needle during surgery, metal pieces leads to reaction in the eye, hence there is harm to the patient.It is concluded as reportable.All available information has been disclosed.The complaint has now been closed.Please note: this incident is being reported as a serious injury since the particles or shavings were seen which reasonably suggests that the particles were retrieved.Metal particles shedding into the eye are likely to result in medical intervention therefore this complaint is being reported as a serious injury.
 
Event or Problem Description
Sales rep reported that during my visit the dr has to use a fragmentation needle to remove a hard lens.To our surprise we saw small pieces of metal in the sclerotomy that came from the needle.Sample is available.All information has been disclosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHACO FRAGMENTATION NEEDLES, 23G / 0.6 MM, INCL. NEEDLE WRENCH
Common Device Name
PHACO FRAGMENTATION NEEDLES, 23G / 0.6 MM, INCL. NEEDLE WRENCH
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7504638
Report Number1222074-2018-00114
Device Sequence Number16173587
Product Code HQC
UDI-Device Identifier08717872016818
UDI-Public08717872016818
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 07/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number3005.F106
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2015
Initial Report FDA Received Date05/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-