BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7578
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the pta balloon dilatation catheter allegedly ruptured at 18 atm.Reportedly, the procedure was completed with another pta balloon catheter.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device was returned for evaluation.The device was attached to an in-house inflation device and was inflated.However, a leak was noted in the balloon.The fibers were stripped from the balloon and a longitudinal rupture was identified.Therefore, the investigation is confirmed for a longitudinal rupture.The definitive root cause for the identified longitudinal rupture could not be determined based upon available information.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiration date: 08/2020).(b)(4).
 
Event or Problem Description
It was reported that the pta balloon dilatation catheter allegedly ruptured at 18 atm.Reportedly, the procedure was completed with another pta balloon catheter.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
parque industrial colonial
tempe AZ 85281
MDR Report Key7505384
Report Number2020394-2018-00649
Device Sequence Number11206493
Product Code DQY
UDI-Device Identifier00801741063015
UDI-Public(01)00801741063015
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7578
Device Catalogue NumberCQ7578
Device Lot NumberREBW0835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Initial Date Received by Manufacturer 04/16/2018
Supplement Date Received by Manufacturer07/27/2018
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date08/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-