| Model Number |
25-322-71-71 |
| Medical Device Problem Codes |
Device Issue (2379); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Impaired Healing (2378)
|
| Date of Event |
03/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Reference exemption number e2017029.Multiple mdr reports were filed for this event, please see associated report: mdr 9610905-2018-00052, mdr 9610905-2018-00054.
|
| |
|
Event or Problem Description
|
|
It was reported that after the patient received radiation the plate may have broken and needs to be removed and replaced.Expected removal date (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed using a stereo microscope which revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.A review of the device history records was not possible due to no lot number identified.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|