| Model Number |
50-347-07-09 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
04/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated report: mdr 9610905-2018-00059 and mdr 9610905-2018-00061.Reference exemption number e2017029.
|
| |
|
Event or Problem Description
|
|
It was reported that screws were found no longer fixated in mandibular bone during the consolidation period of a distraction case.The product was removed.Doctor felt adequate consolidation had occurred post-distraction and no additional fixation methods were used.
|
| |
|
Additional Manufacturer Narrative
|
|
The stereo microscope investigation revealed surface damage but no cracks.Further observation determined there were no indications of material or manufacturing defects observed.The review of the device history records was not possible due to no lot number being identified or provided.The root cause for the failure cannot be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|
Search Alerts/Recalls
|