| Model Number |
29525 |
| Medical Device Problem Code |
Split (2537)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer states a damaged case.Upon inspection, the following issue was found a damaged power cord with exposed copper.No patient involved or harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of one kendall scd 700 system was preformed for the reported condition of, ¿a damaged cased¿.The unit was triaged and the customer¿s reported condition was confirmed and it was found that the power cord was damaged with exposed copper wire.After a visual inspection of the unit under testing, the power cord was observed to be damaged with exposed copper wire.The power cord insulation split exposing copper wiring ands the power cord was replaced.The root cause of the reported symptom was misuse/handling.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
| |
|