MAUDE Adverse Event Report: MDT KYPHON NEUCHATEL MFG KYPHON XPANDER II INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT KYPHON NEUCHATEL MFG KYPHON XPANDER II INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number KX153
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-operative diagnosis: vertebral compression fracture l3 it was reported that during surgery, the balloon ruptured during inflation.The balloon was broken when it was pulled out of the ver tebral body and the surgeon was unable to tell if some of the balloon was left behind in the body.The product came in contact with patient.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual and functional inspection of the ibt assembly confirmed the y adapter has been separated from the outer shaft of the ibt.The balloon of the ibt has also been damaged/ruptured.It appears the balloon ruptured and the shaft was cut from the adapter to help remove the balloon.Unable to identify root cause of damage/failure.Additional ibt was sent back on the pe.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER II INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel 2000
CH  2000
MDR Report Key7506716
Report Number1030489-2018-00666
Device Sequence Number1529168
Product Code HRX
UDI-Device Identifier00643169093652
UDI-Public00643169093652
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K101864
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberKX153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Initial Date Received by Manufacturer 04/12/2018
Supplement Date Received by Manufacturer12/10/2018
Initial Report FDA Received Date05/11/2018
Supplement Report FDA Received Date12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age21 YR
-
-