| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Returned test strips evaluated with defect found.Qc investigation on 5/8/2018 resulted in defect found; returned test strips produce e-7error code.Reported defect not reproduced with returned meter and test strip.Final root cause: rc-016 bent pins on the strip connector caused by the user.Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(4) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.Husband is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 175 and 190 mg/dl.The customer's expected fasting blood glucose test result range is 102 - 119 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 03/31/2019 and open vial date is (b)(6) 2018.Not able to review the meter memory as customer stated that the meter did not turn on.
|
| |
|
Search Alerts/Recalls
|