It was reported that the physician punctured the vessel and inserted the sheath as usual for a picc insertion in the patient's right arm.When he attempted to remove the dilator and wire, he felt a little resistance and found the wire had collapsed.The physician removed the wire slowly and carefully, and confirmed no parts of wire remained in the patient's body, then continued with the procedure.
|
|
Evaluation: a review of the complaint history, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, quality control, and visual inspection of the returned device was conducted during the investigation.The complaint device was returned to cook inc.For evaluation.The wire guide was found to be damaged, elongated, and kinked.The inside diameter of the dilator was found to be within specification.The root cause was found to be likely failure to follow instructions.It was described that the affected access wire guide and dilator were removed together when the event occurred.In the ifu, it states that after the sheath and dilator assembly are introduced over the access wire, the access wire guide is removed and then the longer marked wire guide is inserted.Then, the dilator is removed and the longer, marked, wire guide is left in place to continue the procedure.It is likely that the forces and pressure applied to the access wire guide during removal of the dilator caused the wire to unravel.Use of the more durable, tfe coated, wire guide is recommended and stated in the ifu for use during dilator removal.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|