| Catalog Number |
15858-21 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/20/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The anesthetic is a supplied item and the complaint notification was forwarded to the supplier.Batch record review performed by the supplier for bupivacaine lot # 41393ev revealed no discrepancies that may have contributed to a complaint of this nature.Retained file samples were tested for product potency.The results of potency testing met specification.See attached hospira evaluation report.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
|
| |
|
Event or Problem Description
|
|
It was reported that the marcaine contained in the portex® spinal anesthesia tray was ineffective.No patient death or serious injury was reported in connection with this incident.
|
| |
|
Search Alerts/Recalls
|