| Catalog Number |
A3391-25 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/12/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The anesthetic is a supplied item and the complaint notification was forwarded to the supplier.Batch record review performed by the supplier for bupivacaine lot # 52140ev revealed no discrepancies that may have contributed to a complaint of this nature.All testing was performed and was within specification requirements.Final release potency testing was confirmed at 100.4% of label claim meeting the specification 93.0%-107.0%.Retained samples for the lot in question were visually inspected by (b)(4) qualified personnel and found to be within specification.Additionally smiths medical provided (b)(4) with 10 samples lot # 52140ev from the returned product and (b)(4) visual evaluation found no problem with the ampules.A 15 ampules lot # 52140ev from the returned by the customer product were forwarded by smiths medical to the accredited laboratory for potency testing.The results of the testing met specification 93.0%-107.0%.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
|
| |
|
Event or Problem Description
|
|
It was reported that the bupivacaine contained in the portex® spinal anesthesia tray was ineffective.The patient required general anesthesia as a result of this incident.No patient death or serious injury was reported in connection with this incident.
|
| |
|