• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® SPINAL KITS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® SPINAL KITS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number A4240-25
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/18/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of device history records and incoming records found no discrepancies or anomalies relevant to the complaint.Visual evaluation of one bupivacaine ampule from the returned tray found no visible non-conformities.The drug liquid was found to be clear (colorless) with no particulates visible inside the vial.The anesthetic is a supplied items and the complaint notification was forwarded to the supplier.Records analysis was performed by the supplier for bupivacaine lot #'s 57-237-ev and confirmed that the reported lot continues to meet all specifications.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Event or Problem Description
It was reported that a bupivacaine medication in portex® spinal kits was ineffective and general anesthesia was used.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SPINAL KITS
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7508141
Report Number3012307300-2018-10211
Device Sequence Number1425382
Product Code CAZ
UDI-Device Identifier15019517049879
UDI-Public15019517049879
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue NumberA4240-25
Device Lot Number3205523
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2016
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BUPIVACAINE
-
-