• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® SPINAL KITS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® SPINAL KITS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number NLD-15597-20
Date of Event 10/18/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of device history records and incoming records found no discrepancies or anomalies relevant to the complaint.The anesthetic is a supplied items and the complaint notification was forwarded to the supplier.Records analysis was performed by the supplier for bupivacaine lot #'s 57-237-ev and confirmed that the reported lot continues to meet all specifications.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Event or Problem Description
It was reported that a bupivacaine medication in portex® spinal kits was ineffective.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SPINAL KITS
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7508176
Report Number3012307300-2018-10222
Device Sequence Number1696846
Product Code CAZ
UDI-Device Identifier20351688078688
UDI-Public20351688078688
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue NumberNLD-15597-20
Device Lot Number3199009
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2016
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-