• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® SPINAL KITS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® SPINAL KITS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 49662-25
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
Review of device history records and incoming records found no discrepancies or anomalies relevant to the complaint.Examination of the returned box found 26 sealed original trays lot # 3260417.Visual evaluation of one bupivacaine ampule from each returned tray found no visible non-conformities.The drug liquid was found to be clear (colorless) with no particulates visible inside the vials.The anesthetic is a supplied item and the complaint notification was forwarded to the supplier.Records analysis was performed by the supplier for bupivacaine-marcaine lot #'s 641403a and confirmed that the pfizer conducts the testing prior to lot release and throughout the lifecycle to ensure that the product continues to meet all the registered specifications.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Event Description
It was reported that a marcaine medication in portex® spinal kits was ineffective.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SPINAL KITS
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL NORTH AMERICA
10 bowman drive
keene NH 03431 0724
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key7508194
MDR Text Key108047819
Report Number3012307300-2018-10232
Device Sequence Number1
Product Code CAZ
UDI-Device Identifier00351688051311
UDI-Public00351688051311
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number49662-25
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
MARCAINE
-
-