• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® SPINAL ANESTHESIA TRAYS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® SPINAL ANESTHESIA TRAYS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number A2832-25
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of device history records and incoming records found no discrepancies or anomalies relevant to the complaint.The anesthetic is a supplied item and the complaint notification was forwarded to the supplier.Records analysis was performed by the supplier for bupivacaine-marcaine lot #'s 641403a and confirmed that the pfizer conducts the testing prior to lot release and throughout the lifecycle to ensure that the product continues to meet all the registered specifications.See attached pfizer evaluation report.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Event or Problem Description
It was reported that the bupivacaine contained in the portex® spinal anesthesia tray was ineffective.No patient death or serious injury was reported in connection with this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SPINAL ANESTHESIA TRAYS
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7508211
Report Number3012307300-2018-10242
Device Sequence Number1333695
Product Code CAZ
UDI-Device Identifier00351688061761
UDI-Public00351688061761
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue NumberA2832-25
Device Lot Number3287395
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-