Catalog Number 49661C-25
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of device history records and incoming records found no discrepancies or anomalies relevant to the complaint.Assay testing, performed by the supplier prior to release of bupivacaine-marcaine lot # 680053a, was within specification.Review of the sterilization certificate for the tray lot number 3339098 (po # (b)(4)) revealed that the product was processed according to validated specification requirements and parameters.Smith medical periodically monitors supplied drug components stability post ethylene oxide sterilization throughout their expiry.Our records show that the supplied drugs components continue to meet potency specifications.Complaint history review does not indicate any upward trends for the drug lot numbers, reported in this complaint.The anesthetics are supplied items and the complaint notification was forwarded to the supplier.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Event or Problem Description
It was reported that a bupivacaine medication in portex® spinal kits was ineffective and general anesthesia was used.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SPINAL KITS
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7508232
Report Number3012307300-2018-10255
Device Sequence Number1788615
Product Code CAZ
UDI-Device Identifier00351688047116
UDI-Public00351688047116
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number49661C-25
Device Lot Number3339098
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2017
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BUPIVACAINE
-
-