| Catalog Number |
05.003.126S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reporter¿s complete mailing address was not provided.Reporter¿s phone number was not provided.Reporter¿s contact name was not provided.The manufacturing location is unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that during a total knee arthroplasty surgical procedure, it was observed that the saw blade was broken.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.It was reported that there were no broken parts in the patient's body.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the saw blade was broken off into two parts.It was also determined that the device appeared to have been struck with a very hard metallic object and may have come into contact with a metallic guide.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper handling, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|