Model Number 693575
Medical Device Problem Codes Fracture (1260); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was hearing audible tones.An interrogation of the device was performed and it was found that a lead integrity alert was triggered on the right ventricular (rv) lead due to sic (sensing integrity counter) and hrns (high rate non-sustained) episodes.There were also two alerts triggered for rv lead impedance out of range for high/undefined pacing impedance.It was also noted that potential oversensing was observed on stored egms (electrograms).A possible lead fracture was suspected.The rv lead was explanted and not replaced as the patient received a competitor sub q implantable cardioverter defibrillator (icd).No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7508616
Report Number2649622-2018-08089
Device Sequence Number11311483
Product Code LWS
UDI-Device Identifier00613994446114
UDI-Public00613994446114
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/17/2016
Device Model Number693575
Device Catalogue Number693575
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2018
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DVBB1D1 ICD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age42 YR
-
-