| Model Number |
693575 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was hearing audible tones.An interrogation of the device was performed and it was found that a lead integrity alert was triggered on the right ventricular (rv) lead due to sic (sensing integrity counter) and hrns (high rate non-sustained) episodes.There were also two alerts triggered for rv lead impedance out of range for high/undefined pacing impedance.It was also noted that potential oversensing was observed on stored egms (electrograms).A possible lead fracture was suspected.The rv lead was explanted and not replaced as the patient received a competitor sub q implantable cardioverter defibrillator (icd).No patient complications have been reported as a result of this event.
|
| |
|