• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR TI CANN HINDFOOT ARTHRODESIS NAIL-EX; GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR TI CANN HINDFOOT ARTHRODESIS NAIL-EX; GUIDE Back to Search Results
Catalog Number 03.008.009
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Device is an instrument and is not implanted/explanted.Dhr review was completed for part no.: 03.008.009, lot no.: 1913419: manufacturing location: (b)(4), release to warehouse date: 09.Jul.2008: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.The returned aiming arm was found to have evidence of impact to the locking hole.Additional areas of the device also show evidence of impact marks.The aiming arm device is etched with ¿do not strike this instrument¿ the sleeve was unable to fully pass through the locking hole due to the damage.The sleeve has gouge marks which evidence attempts to force the sleeve through the damaged locking hole.The complaint is considered confirmed and was able to be replicated.The relevant drawings were reviewed during investigation.Further investigation including dimensional inspection and material and hardness reviews are not applicable, as the issue was determined to be related to post manufacturing impact to the aiming arm, which is not intended to be impacted.There were no issues during the manufacture of this product that would contribute to this complaint condition.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent a hindfoot fusion surgery on (b)(6) 2018.The surgeon was using a synthes hindfoot nail.During the surgery, the drill bit sleeve turned out to be stuck in the targeting device while drilling for the locking screws.Since the surgeon exerted some force, the stuck sleeve was removed and the surgery proceeded as scheduled.It was also reported that there was a surgical delay of 10 minutes.The patient outcome was not reported.There is no other information available at this time.During manufacturer¿s investigation of the returned device it was identified that the sleeve was unable to fully pass through the locking hole due to the damage.This condition was re-evaluated and determined to be reportable on (b)(6) 2018.Concomitant device/s reported: unknown locking screw (part # unknown, lot # unknown, qty unknown).This report is for one (1) aiming arm for ti cann hindfoot arthrodesis nail-ex.This is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM FOR TI CANN HINDFOOT ARTHRODESIS NAIL-EX
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7508716
Report Number2939274-2018-52154
Device Sequence Number1164502
Product Code FZX
UDI-Device Identifier10886982066886
UDI-Public(01)10886982066886
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.008.009
Device Lot Number1913419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Initial Date Received by Manufacturer 05/11/2018
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-