Model Number HC105-20
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Codes Corneal Pannus (1447); Calcium Deposits/Calcification (1758); Pulmonary Valve Stenosis (2024)
Date of Event 06/23/2011
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: upon receipt at medtronic's quality laboratory, approximately 3.5cm of the conduit was returned with a transcatheter valve inside.Several pieces of the conduit were returned loose.Pannus fused the devices together, and was also noted on the inner side of the conduit.Calcification was observed.Due to the return condition of the conduit, the leaflets could not be assessed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that eight years post implant of this bioprosthetic valved conduit, implanted in a (b)(6) year old pediatric patient, a transcatheter pulmonary bioprosthetic valve (tpbv) was implanted valve-in-valve.The reason for replacement was reported as moderate to severe stenosis.Six years and nine months post valve-in-valve, both products were explanted and replaced with another bioprosthetic valved conduit.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANCOCK
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7508889
Report Number2025587-2018-01115
Device Sequence Number11053819
Product Code LWR
UDI-Device Identifier00681490908597
UDI-Public00681490908597
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P870078
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2003
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/23/2007
Device Model NumberHC105-20
Device Catalogue NumberHC105-20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2018
Initial Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age17 YR
-
-