MEDTRONIC IRELAND VALIANT CAPTIVIA; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VAMF3434C200TU
Medical Device Problem Code Difficult To Position (1467)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 04/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant captivia stent graft system was planned to be implanted in a patient for the endovascular treatment of an unknown size thoracic aortic aneurysm.It was reported that the physician was unable to pass the valiant captivia stent graft system (b)(4) into the main right femoral artery and was unable to get the device to the desired location.The access vessel size was reported to be 8-9mm and non-calcified.Multiple attempts were made but the graft could not pass.Another manufacturers 24 french sheath was then passed through the access vessel and the vessel dissected.A non-mdt stent graft was implanted to resolve the dissection and the case was aborted.The physician attributed the cause of the event to the patients anatomy and reported that access was extremely difficult.The difficulty with access was not expected due to the size of the access vessels.No additional clinical sequelae were reported and the patient is being monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7510015
Report Number2953200-2018-00696
Device Sequence Number13859020
Product Code MIH
UDI-Device Identifier00643169347366
UDI-Public00643169347366
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2019
Device Model NumberVAMF3434C200TU
Device Catalogue NumberVAMF3434C200TU
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2018
Initial Report FDA Received Date05/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
-
-