(b)(6).Pump passed the displacement test.Unit received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the unexpected restart error test, prime/compromised force sensor system test, excessive no delivery test and occlusion test or verify unexpected restart alarm due to motor error alarm.The test infusion set and reservoir does lock into place.Pump received with minor scratched lcd window, cracked case at the display window corners, cracked lcd window, cracked belt clip slot, cracked battery tube threads, stained end cap sticker, stained address/serial number label and broken reservoir tube lip.
|