ALLERGAN (AUSTIN) STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MX-15-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: deflation- patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
 
Event or Problem Description
Healthcare professional reported a left side deflation.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: brown particles, linear opening, fold creases.A leak test was perform and was found one opening.A microscopic analysis of the returned device identified: a linear sharp opening on posterior side assessed as unidentified(tear) opening(a dimension measurement in the shell was performed which identify the thickness within specification), partial delamination of suture tap assessed as adhesive failure and tree curved sharp openings on suture tap assessed as unidentified(tear) opening.Based on the device analysis the final assessment is: a sharp opening due to unidentified(tear) opening.
 
Event or Problem Description
Healthcare professional reported a left side deflation.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (AUSTIN)
301 w howard lane
suite 100
austin TX 78753
MDR Report Key7510890
Report Number9617229-2018-03527
Device Sequence Number1624025
Product Code LCJ
UDI-Device Identifier10888628004504
UDI-Public10888628004504
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/26/2021
Device Catalogue Number133MX-15-T
Device Lot Number3083627
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Initial Date Received by Manufacturer 04/23/2018
Supplement Date Received by Manufacturer05/29/2018
Initial Report FDA Received Date05/14/2018
Supplement Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-