Model Number N/A |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/04/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the persona implant was opened in the sterile field by the nurse in the surgical room.The surgical tech noticed hair on the plastic packaging.A duplicate item was then opened.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.From the returned product it was determined that foreign material is not found in the package and from the received photos, a hair was noticed inside the cavity.Device history record was reviewed and no discrepancies relevant to the reported event were found.Investigation results concluded that the reported event was due to manufacturing deficiency.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|