• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. VENTRIO ST; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. VENTRIO ST; SURGICAL MESH Back to Search Results
Catalog Number 5950090
Medical Device Problem Codes Defective Device (2588); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Adhesion(s) (1695); Erosion (1750); Unspecified Infection (1930); Pain (1994); Injury (2348); Disability (2371)
Date of Event 10/27/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There is no connection that can be made at this time between the reported unspecified post-operative complications and any problem with the bard/davol ventrio st device used to treat the patient.The patient's attorney did not allege a specific device failure or patient adverse event, and no medical records have been provided.Based on the limited information provided at this time, no conclusions can be made.No lot number has been provided; therefore a review of the manufacturing records is not possible at this time.Should additional information be provided a supplemental emdr will be submitted.Note: the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Not returned.
 
Event or Problem Description
The following was alleged by the patient's attorney: on (b)(6) 2013: the patient underwent surgery for implant of an unspecified bard/davol ventrio st.As reported, the patient is making a claim for an adverse patient outcome against the bard/davol ventrio st.As reported, the attorney alleges general allegations for ¿past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient.¿.
 
Event or Problem Description
The following was alleged by the patient's attorney: (b)(6) 2013: the patient underwent surgery for implant of an unspecified bard/davol ventrio st.As reported, the patient is making a claim for an adverse patient outcome against the bard/davol ventrio st.As reported, the attorney alleges general allegations for ¿past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient.¿ addendum per additional information provided: (b)(6) 2013 - patient was diagnosed with ventral incisional hernia thereby underwent open repair with the implant of ventrio st.Per operative notes, ¿moderate amount of adhesions were noted upon entering the abdomen and were taken down.A ventrio st was placed using sutures.¿ (b)(6) 2015 - patient was diagnosed with erosion of prosthetic mesh in the colon thereby underwent removal of infected mesh and implant of biologic mesh.Per operative notes, ¿there were extensive lysis of adhesions.The omentum was stuck to the posterior aspect of the mesh.The mesh had obviously eroded into the left colon.The mesh was removed.A biologic mesh was then placed.¿ attorney alleged that the patient had adhesions, bowel obstruction, bowel perforation, bowel removal, infection, mesh migration, pain, hernia recurrence and emotional injuries.
 
Additional Manufacturer Narrative
There is no connection that can be made at this time between the reported unspecified post-operative complications and any problem with the bard/davol ventrio st device used to treat the patient.The patient's attorney did not allege a specific device failure or patient adverse event, and no medical records have been provided.Based on the limited information provided at this time, no conclusions can be made.No lot number has been provided; therefore a review of the manufacturing records is not possible at this time.Addendum: this supplemental emdr is submitted to document additional information provided: based on the additional information received, there is no change to the initial determination, no conclusions can be made.Per medical records review, about 2 years 6 months post implant of ventrio st, patient was diagnosed with infection, mesh erosion and adhesions thereby underwent repair with mesh removal.The instructions-for-use supplied with the device list infection and erosion as possible complications.In regard to infection, the warnings section in ifu states, "if an infection develops, treat the infection aggressively.Consideration should be given regarding the need to remove the mesh.An unresolved infection may require the removal of the mesh." a review of the manufacturing records was performed and found that the lot was manufactured to specification.Note, the manufacturing date (15-may-2011) is considered to be a best estimate.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRIO ST
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key7512443
Report Number1213643-2018-01488
Device Sequence Number1624037
Product Code FTL
UDI-Device Identifier00801741031588
UDI-Public(01)00801741031588(17)160528(10)HUVE0967
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K101920
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2015
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/28/2016
Device Catalogue Number5950090
Device Lot NumberHUVE0967
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2018
Supplement Date Received by Manufacturer01/02/2025
Initial Report FDA Received Date05/14/2018
Supplement Report FDA Received Date01/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability; Required Intervention;
Patient Age38 YR
Patient SexMale
-
-