• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INC. ALIF INTERFIXATED SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INC. ALIF INTERFIXATED SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR Back to Search Results
Model Number 6955202
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned for evaluation as it remains in-situ.A review of the x-ray pictures provided confirm the alleged event.No revision procedure is planned at this time.Labeling: potential adverse events and complications ".As with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)." patient education: ".The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." product remains in-situ.
 
Event or Problem Description
On (b)(6) 2017 patient underwent a procedure using the base interfixated titanium system.On (b)(6) 2018 during the six month follow up, a screw fracture was noted.Patient is asymptomatic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALIF INTERFIXATED SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
MDR Report Key7512606
Report Number2031966-2018-00057
Device Sequence Number10831606
Product Code OVD
UDI-Device Identifier00887517701749
UDI-Public00887517701749
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K151214
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number6955202
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2018
Initial Report FDA Received Date05/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-