Consumer reported complaint for high blood glucose results.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 115 -116 mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the bathroom.The test strip manufacturer's expiration date is 07/25/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1:328 mg/dl, date:(b)(6) 2018, time:02:14 pm, fasting; result 2:232 mg/dl, date:(b)(6) 2018, time:03:49 pm, fasting; result 3:241 mg/dl, date:(b)(6) 2018, time:11:02 pm, fasting; result 4:143 mg/dl, date:(b)(6) 2018, time:08:52 am, fasting; result 5:249 mg/dl, date:(b)(6) 2018, time:09:20 pm, fasting.The customer is stating that when he compares the results from his meter to his physician's meter (at scheduled routine visit) it is reading 100 points difference.
|