| Model Number |
MMT-723LNAS |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump alarmed motor error.The customer¿s blood glucose level was 130 mg/dl at the time of the incident.The customer stated that the drive support cap was normal and that the insulin pump was not exposed to high magnetic field and was able to complete the rewind process.Customer reported a motor position encoder error.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The insulin pump passed rewind test, basic occlusion test, prime alarm, compromised force sensor test and displacement test.However, the insulin pump received stuck in the motor error loop during bolus/basal delivery and motor position encoder alarm confirmed in the history file.The motor was tested outside the device and passed.The motor may have had an intermittent failure that was not detected during our testing.
|
| |
|
Search Alerts/Recalls
|