MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 GLOBAL SHD END PEG GLEN 48; GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 GLOBAL SHD END PEG GLEN 48; GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID Back to Search Results
Catalog Number 113786025
Medical Device Problem Code Failure To Adhere Or Bond (1031)
Health Effect - Clinical Codes Not Applicable (3189); No Code Available (3191)
Date of Event 04/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states that the patient's rotator cuff failed so the surgeon pulled everything out and converted to a djo reverse.The patient was revised to address loosening of the glenoid component at the bone to cement and cement to implant interface.Depuy cement was used.Doi: unknown.Dor: (b)(6) 2018.Left shoulder.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL SHD END PEG GLEN 48
Common Device Name
GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7513793
Report Number1818910-2018-59595
Device Sequence Number19298379
Product Code KWT
UDI-Device Identifier10603295007265
UDI-Public10603295007265
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K905786
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number113786025
Device Lot NumberW24JW1002
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/18/2018
Supplement Date Received by Manufacturer11/21/2018
Initial Report FDA Received Date05/15/2018
Supplement Report FDA Received Date11/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-