| Model Number |
H7493952824300 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).Device evaluated by mfr.: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed. (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.00x24mm promus premier¿ drug-eluting stent was selected for use to treat the lesion.However, it was noticed that the tip of the stent seemed to be damaged and the struts were not in placed.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: promus premier us mr 3.00x24mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The photo attached is one image of the device that represents the observation of no damage.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of shaft polymer extrusion revealed no issues.A visual and microscopic examination found no damage to the tip.No issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.00x24mm promus premier drug-eluting stent was selected for use to treat the lesion.However, it was noticed that the tip of the stent seemed to be damaged and the struts were not in placed.The procedure was completed with a different device.No patient complications were reported.
|
| |
|