• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE EXPRESS® LD ILIAC / BILIARY; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE EXPRESS® LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number H74938047720130
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that stent dislodgment occurred.The target lesion was located in the iliac vessel.After placing a 7f guide catheter ostium of iliac, a 7.0x20x135 cm express® ld iliac / biliary stent was advanced through the guide catheter forcibly.However, upon pulling the device back, the stent sheared off from the balloon into the abdominal aorta.Multiple attempts were made to remove the stent and finally was able to snare the stent out of the sheath.The procedure was completed with a 8f guide catheter and a different stent was used.No patient complications were reported and the patient was not harmed.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the express ld delivery system was returned for analysis with the stent detached from the balloon material.The stent was not returned for analysis.One kink was noted on the returned device.A visual and microscopic examination was performed on the returned device, the following attributes were examined: the delivery system was returned for analysis.A visual and microscopic examination was performed on the returned delivery system.One kink was identified on the shaft of the device approximately 11mm distal to the proximal markerband within the balloon material.This type of damage is consistent with excessive force being applied to the delivery system.The balloon material was slightly unfolded and appeared to have been subjected to positive pressure.This may have occurred during attempts the remove the device.No issues were noted with the returned device that could have contributed to the complaint incident.The investigation conclusion is user / use error as there was an act or omission of an act that resulted in a different medical device response than intended by the manufacturer or expected by the user.The dfu states: the recommended introducer sheath size for this express ld device is a 6fr introducer sheath.If strong resistance is met during advancement of the premounted stent system, discontinue movement and if the cause of resistance cannot be determined, withdraw both the premounted stent system and sheath or guide catheter as a single unit.(b)(4).
 
Event or Problem Description
It was reported that stent dislodgment occurred.The target lesion was located in the iliac vessel.After placing a 7f guide catheter ostium of iliac, a 7.0x20x135 cm express ld iliac/biliary stent was advanced through the guide catheter forcibly.However, upon pulling the device back, the stent sheared off from the balloon into the abdominal aorta.Multiple attempts were made to remove the stent and finally was able to snare the stent out of the sheath.The procedure was completed with a 8f guide catheter and a different stent was used.No patient complications were reported and the patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS® LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7513849
Report Number2134265-2018-04125
Device Sequence Number19258378
Product Code NIO
UDI-Device Identifier08714729392439
UDI-Public08714729392439
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P090003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/23/2020
Device Model NumberH74938047720130
Device Catalogue Number38047-72013
Device Lot Number0021429962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Initial Date Received by Manufacturer 04/20/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date05/15/2018
Supplement Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-