| Model Number |
M00572230 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.However, the complainant stated that the device was used prior to the expiration date.(b)(4).The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018, that an epic biliary stent had been implanted to treat a malignant stricture in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was tortuous and was dilated prior to stent placement.According to the complainant, during the procedure, the stent was able to be deployed; however, the delivery system got caught on the stent upon removal.It was noted that the stent slightly moved but was still in a good location.The delivery system was then successfully removed.The stent remained implanted inside the patient and the procedure was completed at this time.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
| |
|