• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS MESH SOFRADIM - PARIETEX¿ OPTIMIZED COMPOSITE MESH; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS MESH SOFRADIM - PARIETEX¿ OPTIMIZED COMPOSITE MESH; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number PCO8OSX
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Hematoma (1884); Inflammation (1932); Pain (1994); Scar Tissue (2060); Hernia (2240); Injury (2348); Obstruction/Occlusion (2422); No Code Available (3191); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment.It was reported that after a procedure where this device was implanted, the patient experienced recurrence and revision.After the original procedure, additional potentially related procedures were performed, including revision surgery.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an umbilical hernia.It was reported that after a procedure where this device was implanted, the patient experienced recurrence.Treatment included revision of the mesh five years after implant.During the revision, scar tissue was noted.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic. the fda was notified of this large complaint receipt. due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an umbilical hernia.It was reported that after a procedure where this device was implanted, the patient experienced recurrence, adhesions, obstruction, pain, hematoma, inflammation.Treatment included medication, partial revision, hernia repair with new mesh, revision of the mesh five years after implant.During the revision, scar tissue was noted.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an umbilical hernia.It was reported that after a procedure where this device was implanted, the patient experienced recurrence, adhesions, obstruction, pain, scar tissue, hematoma, inflammation.Treatment included medication, partial revision, hernia repair with new mesh, revision of the mesh five years after implant.
 
Additional Manufacturer Narrative
Additional information : (patient codes).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an umbilical hernia.It was reported that after a procedure where this device was implanted, the patient experienced recurrence, adhesions, obstruction, pain, scar tissue, hematoma, inflammation, <(>&<)> edematous omentum.Post operative treatment included medication, partial revision, hernia repair with new mesh, revision of the mesh, partial mesh removal, hernia repair with sutures, <(>&<)> scar tissue <(>&<)> mesh excised.
 
Additional Manufacturer Narrative
Additional info: a4, b5, h6 (patient codes), <(>&<)> additional codes medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: a4, d4 (udi), g1.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESH SOFRADIM - PARIETEX¿ OPTIMIZED COMPOSITE MESH
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
mike bocchino
8200 coral sea st ne
8200 coral sea st ne
mounds view,mn, CT 55112-4391
8006464633
MDR Report Key7514175
Report Number9615742-2018-01150
Device Sequence Number1760034
Product Code FTL
UDI-Device Identifier10884521179721
UDI-Public(01)10884521179721(17)131130(10)PLL00034
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K110816
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2016
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 08/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2013
Device Model NumberPCO8OSX
Device Catalogue NumberPCO8OSX
Device Lot NumberPLL00034
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/18/2018
Supplement Date Received by Manufacturer06/20/2018
08/08/2018
05/15/2019
12/06/2023
12/20/2023
03/26/2025
07/30/2025
Initial Report FDA Received Date05/15/2018
Supplement Report FDA Received Date07/16/2018
10/03/2018
10/30/2019
12/12/2023
12/22/2023
04/03/2025
08/01/2025
Date Device Manufactured12/12/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age57 YR
Patient SexMale
Patient Weight113 KG
-
-