| Catalog Number |
136352000 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Not Applicable (3189)
|
| Date of Event |
10/24/2012
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Pinnacle ppf and sticker sheets received.Ppf alleges dislocation, cup and stem loosening.It was also indicated in stickers sheets that tapered spacer articuleze, modular cathcart ball and summit por taper were implanted on (b)(6) 2011.Doi: (b)(6) 2011; dor: (b)(6) 2012; left hip.(b)(4) - 1st revision ;(b)(4) -2nd revision; (b)(4) -3rd revision.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|