| Model Number |
ADDR06 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Muscle Stimulation (1412)
|
| Date of Event |
04/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 507652 lead, implanted: (b)(6) 2009.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced pocket stimulation after a device change out.It was also noted that the device was implanted sub-muscular which would likely have been the cause of the new pocket stimulation.Reprogramming was performed which significantly decreased the stimulation.The implantable pulse generator (ipg) remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|