Model Number X SERIES
Medical Device Problem Codes Delayed Charge Time (2586); Device Operational Issue (2914)
Health Effect - Clinical Code Death (1802)
Date of Event 03/17/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while attempting to monitor a (b)(6) male patient, the device delayed to charge and internally dumped the energy after charging up to defibrillate the patient.Complainant indicated that the patient subsequently expired, however they believe it was not a result of the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the customer's report for device slow to charge was observed during review of the device's activity log.However, the reported problem could not be duplicated with the device.The battery used during the event was not returned as part of the investigation.The battery pin connection hardware was replaced as a precaution.The device was recertified and returned to the customer.Review of the activity logs for the customer's report the device disarmed itself was confirmed ; however the device performed to specification.The device is designed to internally discharge once a charge attempt exceeds 25 seconds.This occurred on both second and third charge attempts since the charge time exceeded 25 seconds.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7514637
Report Number1220908-2018-01329
Device Sequence Number1529280
Product Code MKJ
UDI-Device Identifier00847946004330
UDI-Public00847946004330
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K112432/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/24/2018
Supplement Date Received by Manufacturer04/24/2018
Initial Report FDA Received Date05/15/2018
Supplement Report FDA Received Date06/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age35 YR
-
-