MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC ELITE COMPRESSION IMPLANT KIT 15X15X15MM 2 LEGS; STAPLE,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC ELITE COMPRESSION IMPLANT KIT 15X15X15MM 2 LEGS; STAPLE,FIXATION,BONE Back to Search Results
Model Number EL-1515S2
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information is not available for reporting.Due to the intra-operative events, the device was not successfully implanted.An alternate device was used to complete procedural step.As such, implant/explant dates are not applicable.Complainant device is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the surgeon was attempting to implant one of the bme elite staples in proximal tibia when the staple would not release from the handle (the release trigger would not move).The surgeon was pulling with two hands to release the trigger and it would not release.The sales consultant opened a second implant and it worked perfectly.The surgery continued without incident and there was not patient consequence.This is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELITE COMPRESSION IMPLANT KIT 15X15X15MM 2 LEGS
Common Device Name
STAPLE,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7514889
Report Number2939274-2018-52182
Device Sequence Number1225226
Product Code JDR
UDI-Device Identifier00810633021675
UDI-Public(01)00810633021675
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K150125
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEL-1515S2
Device Catalogue NumberEL-1515S2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/24/2018
Initial Report FDA Received Date05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-