• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC WIDEBAND I 30BX; MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC WIDEBAND I 30BX; MALE EXTERNAL CATHETER Back to Search Results
Model Number 36303
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the male external catheter's adhesive was strong, as a result it was difficult to remove the catheter.It was reported that a warm compress was used to try and remove the male external catheter, however the catheter was still not coming off.Allegedly it took 30 minutes to remove the male external catheter and scissors had to be used to remove the catheter.Allegedly the patient only changed on the male external 2 times a week, when the instructions state to change the catheter every 24 hours.
 
Additional Manufacturer Narrative
The reported event was unconfirmed since the problem could not be reproduced.Per visual inspection, the sample was compared to a production sample and no differences were noted.The sample was evaluated according to the alleged failure mode: "strong adhesive".The evaluation was performed and the tested sample was found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: " the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove: gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." correction: mfr site.
 
Event or Problem Description
It was reported that the male external catheter's adhesive was strong, as a result it was difficult to remove the catheter.It was reported that a warm compress was used to try and remove the male external catheter, however the catheter was still not coming off.Allegedly it took 30 minutes to remove the male external catheter and scissors had to be used to remove the catheter.Allegedly the patient only changed on the male external 2 times a week, when the instructions state to change the catheter every 24 hours.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC WIDEBAND I 30BX
Common Device Name
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7515115
Report Number1018233-2018-01745
Device Sequence Number1386024
Product Code NNX
UDI-Device Identifier00801741071065
UDI-Public(01)00801741071065
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2019
Device Model Number36303
Device Catalogue Number36303
Device Lot Number53426774
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2018
Initial Date Received by Manufacturer 04/25/2018
Supplement Date Received by Manufacturer07/27/2018
Initial Report FDA Received Date05/15/2018
Supplement Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-