Investigation ¿ evaluation: the device was not returned and no photographs were provided.Without the complaint device, a physical investigation was not able to be completed.A document based investigation was performed.A review of complaint history, the device history record, instructions for use, manufacturing instructions and quality control data was conducted.The device history record was reviewed, for lot 8078345 (finished assembly), lot sa8003365 (sheath assembly), and sa7818503/sa7827234/sa7908770 (captor valve assembly).There were non-conformances identified on sa7818503, sa7827234 and sa798770 and all non-conforming items were scrapped, giving no indication of additional non-conforming product.A review of complaint history records revealed this complaint to be the only reported complaint associated to the complaint lot number 8078345.The common causes of valve leakages include damage/tearing of silicone discs, aggressive handling of the valve, or inserting the dilator through the valve when it is in a locked position.Other possible causes could be that the silicone discs were not assembled correctly or the iris valve was not functioning properly due to insufficient or excessive glue applied.Also, improper storage conditions or damage during shipping/handling cannot be ruled out as possible causes for the leakage.Without the return of the complaint device, a definitive cause for this complaint cannot be determined.No further risk reduction activities will be required at this time.Cook medical has notified the appropriate personnel and the will continue to monitor this device via the complaints database for similar complaints.
|
|
It was reported, during preparation, it was unable to remove air from the delivery system by flushing because the flushing fluid leaked from the hemostatic valve.Another zenith flex with spiral-z technology aaa endovascular graft iliac leg was used to complete the procedure.No patient impact was reported as the malfunction was discovered prior to patient contact.
|