LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 04/16/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message on the patient side during maintenance.There was no patient involvement.
 
Manufacturer Narrative
Through follow up communication livanova (b)(4) learned that the device was repaired by a third party provider.However it was not reported what part was found to be the root cause.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial.
 
Manufacturer Narrative
Through follow up communication, livanova deutschland learned that the service activity was performed by a service provider and that the issue was solved by the replacement of the failing patient pump.The affected pump was requested back to the manufacturer site for a detailed investigation.Results revealed that the pump was faultless and working within specifications.The involvement of the part in the reported failure could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key7516136
MDR Text Key108636394
Report Number9611109-2018-00959
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 04/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received03/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-