| Model Number |
EGIAUSTND |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Codes |
Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during low anterior resection, the surgeon fully fired the device, however they could not pull the black return knob back to the initial position and could not open the jaws.The surgeon additionally resected the tissues of the mouth side and anus side, and removed the cartridge from the tissue, then sutured the resected part manually.The surgical time was extended more than 30 minutes.The incision site was not extended.Nothing fell into the patient's cavity.No bleeding occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received engaged with the subject reload.The subject reload was unloaded without difficulty.Initial visual inspection of the instrument noted no abnormalities.Functionally, the instrument was loaded with an articulating vascular/medium reload.During the firing cycle, a skip was audible in the firing stroke.An access hole was cut into the instrument body for visualization of the firing rack.This examination noted sheared teeth on the firing rack.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the sheared teeth on the firing rack may occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and precaution."1.Preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in c or responding selection of staple size.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|