MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO SENSIA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO SENSIA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number SEDR01
Medical Device Problem Codes Failure to Interrogate (1332); Pacing Problem (1439)
Health Effect - Clinical Code Bradycardia (1751)
Date of Event 04/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented to the emergency room (er) with a bradycardic heart rate and no pacemaker spikes were visible on the external monitor.The implantable pulse generator (ipg) was unable to connect with the programmer via telemetry.A magnet was placed over the ipg and there was no discernible pacing output.The device had reached end of life (eol).The patient had been lost-to-follow-up and was last seen three to five years ago, the patient did not recall the exact timeframe.The device was extracted and replaced.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIA DR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7516342
Report Number3004209178-2018-10981
Device Sequence Number13879028
Product Code NVZ
UDI-Device Identifier00613994177957
UDI-Public00613994177957
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2009
Device Model NumberSEDR01
Device Catalogue NumberSEDR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2018
Initial Date Received by Manufacturer 04/24/2018
Supplement Date Received by Manufacturer07/26/2018
08/08/2018
Initial Report FDA Received Date05/15/2018
Supplement Report FDA Received Date07/26/2018
10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5054-52, LEAD / 5568-45, LEAD
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age70 YR
Patient Weight92
-
-