| Catalog Number |
1272.EDPT5 |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/13/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation showed that the lime stone effect happened due to the cyanoacrylate glue.This is discussed with the production department and the procedure has been changed.Cannula seems to have trace of " limestone" hence concluded as reportable.The associated risk is foreign body reaction.(b)(4).All available information has been disclosed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
We received a complaint it seems to have trace of " limestone" on it.Sample available.
|
| |
|