• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC SV CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET FULL; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC SV CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET FULL; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAY Back to Search Results
Model Number 1274108
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a tear in the extension leg near the hub was confirmed, but the exact cause of the damage was unknown.One 4fr d/l powerpicc sv was returned for investigation.The catheter exhibited evidence of use.A circumferential breach was observed in the purple extension leg just inside the distal end of the red connector.The cross section of the damaged extension leg was microscopically examined and was noted to be rough and jagged.The characteristics observed on the extension leg indicate that the tubing tore.The break in the extension leg could be attributable to pulling, twisting, and manipulating the luer adaptor.An annular wear pattern was observed in the extension leg that coincided with the distal end of the red connector.A semi-circumferential crease was observed 3mm distal to the location of the breach.The crease in the tubing indicates that the extension leg was kinked at that location.Due to the evidence of use, the damage may have developed over time.It was reported that the picc was placed (b)(6) 2017 and removed on (b)(6) 2018.No further action is required because the reported event could not be related to a deficiency in product manufacture or deficiency while under correct product use.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that the catheter was removed due to break/separation where red hub connects to the extension leg.No injury was reported and a new line was placed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SV CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET FULL
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAY
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7516921
Report Number3006260740-2018-00963
Device Sequence Number1544933
Product Code LJS
UDI-Device Identifier00801741086908
UDI-Public(01)00801741086908
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1274108
Device Catalogue Number1274108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2018
Initial Report FDA Received Date05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-