| Model Number |
800SC26 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Mitral Regurgitation (1964)
|
| Date of Event |
04/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of this 26mm annuloplasty band in the mitral position, the band was explanted and replaced with a 34mm annuloplasty band due to residual regurgitation.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: event appears to be sizing related as the physician opted to use a larger band.If information is provided in the future, a supplemental report will be issued.
|
| |
|