Catalog Number K09A
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor applicable imaging films were returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented with pre-op diagnosis, primary osteoporosis and compression fracture.Intra-op, the inflatable bone tamp failed during the first insertion attempt into the vertebral body.When pressure was applied to one place while inflating the balloon, it deformed.The balloon was put out of the body once and inflated, it did not inflate as a whole and one place of the balloon inflated like baking a rice cake.A 1 ml of the contrast medium was added to the balloon but the balloon did not inflate (pressure 300psi).The contrast medium was added more and when 1.5 ml was put into the balloon, the balloon deformed in the vertebral body and only a part of it inflated (pressure 380psi).The device came in contact with the patient.No patient complications were reported as a result of the event.After occurrence of the event, a new kit was opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38118
MDR Report Key7517492
Report Number1030489-2018-00688
Device Sequence Number16500257
Product Code HRX
UDI-Device Identifier00858196001161
UDI-Public00858196001161
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK09A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2018
Initial Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-