• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE Back to Search Results
Catalog Number 8268.VIT25
Medical Device Problem Code Device Operational Issue (2914)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/21/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation showed that the part that connects the tubing to the vitrectome was broken and the outer knife was in a angle.The investigation regarding the air bubbles could not be done due to the broken part.Assuming worst case during surgery, air bubbles observed at the probe.This may cause injury to the patient.Hence it is reportable.The risk is possible injury by microbial, viral or prion infection (patient harm) this reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4)).Details of this activity were discussed with cdrh office of compliance ((b)(6)) during a tele-conference on (b)(4) 2017.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
During operation, air bubble happened.Situation is as follows; eva shaving mode, 8000cpm, flow:ss, vaccume 600.Please investigate it carefully and find root cause, and improve.Sample is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE
Common Device Name
DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7518018
Report Number1222074-2018-00134
Device Sequence Number16526065
Product Code HQE
UDI-Device Identifier08717872020808
UDI-Public08717872020808
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 10/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number8268.VIT25
Device Lot Number2000371167
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2015
Initial Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-